Number of channels | R&SRTE1022/R&SRTE1032/ R&SRTE1052/R&SRTE1102/ R&SRTE1152/R&SRTE1202 | 2 |
| R&SRTE1024/R&SRTE1034/ R&SRTE1054/R&SRTE1104/ R&SRTE1154/R&SRTE1204 | 4 |
Analog bandwidth (–3 dB) at 50 Ω | R&SRTE1022/R&SRTE1024 | ≥ 200 MHz |
| R&SRTE1032/R&SRTE1034 | ≥ 350 MHz |
| R&SRTE1052/R&SRTE1054 | ≥ 500 MHz |
| R&SRTE1102/R&SRTE1104 | ≥ 1 GHz |
| R&SRTE1152/R&SRTE1154 | ≥ 1.5 GHz |
| R&SRTE1202/R&SRTE1204 | ≥ 2 GHz |
Rise time | R&SRTE1022/R&SRTE1024 | < 1.75 ns |
| R&SRTE1032/R&SRTE1034 | < 1 ns |
| R&SRTE1052/R&SRTE1054 | < 700 ps |
| R&SRTE1102/R&SRTE1104 | < 350 ps |
| R&SRTE1152/R&SRTE1154 | < 233 ps |
| R&SRTE1202/R&SRTE1204 | < 175 ps |
Impedance | | 50 Ω ± 1,5 %, 1 MΩ ± 1 % || 16 pF ± 1 pF (meas.) |
Input sensitivity | max. bandwidth in all ranges | 50 Ω: 500 μV/div to 1 V/div 1 MΩ: 500 μV/div to 10 V/div |
Resolution | | 8 bit (7-bit ENOB) |
| with high definition mode | 16 bit |
Acquisition system |
Realtime sampling rate | | max. 5 Gsample/s on each channel |
Acquisition memory | standard configuration, per channel/1 channel active | R&SRTE1000 2-channel model: 10/20 Msample R&S&rE1000 4-channel model: 10/40 Msample |
| max. upgrade (R&SRTE-B102 option), per channel/1 channel active | R&SRTE1000 2-channel model: 50/100 Msample R&S&rE1000 4-channel model: 50/200 Msample |
Acquisition rate | | > 1 000 000 waveforms/s |
Decimation modes | any combination of decimation mode and waveform arithmetics | sample, peak detect, high resolution, root mean square |
Waveform arithmetics | | off, envelope, average |
Horizontal system |
Timebase range | | 50 ps/div to 5000 s/div |
Timebase accuracy | after delivery/calibration | ±2 ppm |
Channel deskew | | ±100 ns |
Trigger system |
Trigger types | | edge, glitch, width, runt, window, timeout, interval, slew rate, data2clock, pattern, state, serial pattern, TV/video, serial bus trigger (optional) |
Sensitivity | definition of trigger hysteresis | can be set automatically or manually from 0 div to 5 div |
Analysis and measurement functions |
Automated measurements | | 77 measurement functions |
Cursor measurements | | 2 cursor sets, each consisting of two horizontal and two vertical cursors |
Waveform mathematics | | 4 math waveforms; mathematics, logical operations, comparison, FIR filter, FFT |
MSO option |
Digital channels | | 16 (2 logic probes) |
Input impedance | | 100 k || 4 pF |
Sampling rate | | 5 Gsample/s per channel |
Acquisition memory | | 100 Msample per channel |
Parallel buses | | up to 4 |
General data |
Dimensions | W x H x D | 427 mm × 249 mm × 204 mm (16.81 in × 9.8 in × 8.03 in) |
Weight | without options, nominal | 8,6 kg (18.96 lb) |
Screen | | 10.4" LC TFT color touchscreen, 1024 × 728 pixel (XGA) |
Interfaces | | 1 Gbit/s LAN, 4 × USB 2.0, GPIB (optional), DVI for external monitor, external trigger, trigger output |