SPECIFICATIONS |
MODEL |
7450 |
7451 |
7452 |
AC WITHSTAND VOLTAGE |
Output Rating |
5KVAC/100mA Constant 500VA |
Range |
Resolution |
Accuracy |
Output Voltage, ACV |
0 - 5000 |
1 |
±(2% of setting + 5V) |
Output Frequency |
50Hz/60Hz ±100ppm, User Selection |
Output Waveform |
Sine Wave ,THD.<2% (Resistive Load), Crest Factor = 1.3 - 1.5 |
Output Regulation |
± (1% of output + 5V), From no load to full load |
SETTINGS |
Hi-Limit AC Current, mA |
0 - 99.99 |
0.01 |
±(2% of setting + 3counts) |
Lo-Limit AC Current, mA |
0 - 9.999 |
0.001 |
±(2% of setting + 3counts) |
Ramp Time, second |
0.1 - 999.9 |
0.1 |
±(0.1% + 0.05sec) |
Dwell Time, second |
0, 0.3 - 999.9 (0=continuous) |
0.1 |
Arc Detection |
1 - 9 ranges (9 is the most sensitivity) |
AC Current Offset, mA |
0 - 2.000 or Auto Set |
DC WITHSTAND VOLTAGE (7451,7452 ONLY) |
Output Rating |
6KVDC/9999μA |
Output Voltage, DCV |
0 - 6000 |
1 |
±(2% of setting + 5V) |
Output Ripple |
<5%(6KV/9999μA at Resistive Load) |
SETTINGS |
Hi-Limit DC Current, μA |
0 - 9999 |
1 |
±(2% of setting + 3counts) |
Lo-Limit DC Current, μA |
0 - 999.9 |
0.1 |
±(2% of setting + 3counts) |
Ramp Time, second |
0.1 - 999.9 |
0.1 |
±(0.1% + 0.05sec) |
Dwell Time, second |
0, 0.4 - 999.9 (0=continuous) |
0.1 |
Arc Detection |
1 - 9 ranges (9 is the most sensitivity) |
DC Current Offset, μA |
0 - 200.0 or Auto Set |
Ramp High Current |
ON/OFF, User Selection |
Charge - Low Current, μA |
0 - 350.0 or Auto Set |
Discharge Time |
< 200 msec |
INSULATION RESISTANCE |
Output Rating |
1KVDC/9999MΩ |
Output Voltage, DCV |
100 - 1000 |
1 |
±(2% of setting + 3V) |
Hi-Limit Resistance, MΩ |
0, 1 - 9999 (0=OFF) |
1 |
±(2% of setting+2counts)at 500-1000VDC, 1 -1000 MΩ ±(8% of setting+2counts)at 500-1000VDC, 1000 -9999 MΩ ±(8% of setting+2counts)at 100-500VDC, 1 -1000 MΩ |
Lo-Limit Resistance, MΩ |
1 - 9999 |
1 |
Delay Time, second |
0, 0.5 - 999.9 (0=continuous) |
0.1 |
±(0.1% + 0.05sec) |
Charge - Low Current, ?A |
0 - 3.500 or Auto Set |
GROUND BOND (7452 ONLY) |
Output Rating |
30A/600mΩ, 8V |
Output AC Current, A |
3.00 - 30.00 |
0.01 |
± (2% of setting +2 counts) |
Output AC Voltage, V |
3.00 - 8.00 |
0.01 |
± (2% of setting +3 counts) |
Output Frequency |
50Hz/60Hz ±100ppm, User Selection |
SETTINGS |
Hi-Limit Resistance, mΩ |
0 - 150 (3 - 30A) |
1 |
± (2% of setting +2 counts) |
Lo-Limit Resistance, mΩ |
0 - 600 (3 - 10A) |
Lead Resistance Offset, mΩ |
0 -200 |
1 |
± (2% of setting +2 counts) |
Dwell Time, second |
0, 0.5 - 999.9 sec (0 = continuous) |
0.1 |
±(0.1% + 0.05sec) |
MEASUREMENT |
AC/DC Voltage, KV |
0 - 6.00 |
0.01 |
±(1.5% of reading + 1count) |
DC Voltage, V (IR) |
100 - 1000 |
1 |
±(1.5% of reading + 3count) |
AC CurrentRange, mA |
0-3.500 |
0.001 |
±(2% of reading + 3counts) |
3.00-99.99 |
0.01 |
±(2% of reading + 6counts) |
DC Current Range, μA |
0-350.0 |
0.1 |
± (2% of reading + 3counts) |
300 - 3500 |
1 |
± (2% of reading + 2counts) |
3000-9999 |
10 |
± (2% of reading + 6counts) |
AC Current, A (GB) |
0 - 35.00 |
0.01 |
±(2% of reading+2counts) |
Resistance, , MΩ(IR) |
1 - 9999 (AutoRange) |
— |
±(2% of reading+2counts)at 500-1000VDC, 1 -1000 MΩ |
±(8% of reading+2counts)at 500-1000VDC, 1000 -9999 MΩ |
±(8% of reading+2counts)at 100-500VDC, 1 -1000 MΩ |
Resistance, , mΩ(GB) |
0 - 600 |
1 |
±(2% of reading+2counts) |
GENERAL |
Input Voltage AC |
100/115/200/230Vac±15%, 50/60Hz ± 5%, Fuse 15A |
PLC Remote Control |
Input : Test , Reset , Memory1.2.3 |
Output : Pass , Fail , Processing |
Memory |
50 memories, 8 steps/memory |
Display |
20×2 LCD with back light |
Key Lock |
To prevent unauthorized alteration of the test Parameters |
Calibration |
Build-in software and external calibrated meters |
Alarm Volume Setting |
Range: 0-9 ;0=OFF, 1 is softest volume, 9 is loudest volume. |
Rear Panel Output |
YES |
YES |
YES |
Built-in Scanner |
- |
- |
- |
External Scanner |
- |
YES |
YES |
Smart GFI |
YES |
YES |
YES |
Analog Meter |
YES |
YES |
YES |
RS232 Interface |
- |
Option |
Option |
GPIB Interface |
- |
YES |
YES |
Printer Interface |
- |
Option |
Option |
TUV-GS listed |
YES |
- |
YES |
CE Compliance |
YES |
- |
YES |
Environment |
0-40℃, 20-80%RH |
Dimension (W×H×D) |
430×133×400 (mm) |
430×133×400 (mm) |
430×133×500 (mm) |
Net Weight |
28Kg |
28Kg |
32Kg |
STANDARD ACCESSORIES |
Power Cord(15A) |
×1 |
Fuses |
×2 (Including a spare contained in the fuse holder) |
Hipot Test Lead ,1.5m(1101) |
×1 |
×1 |
×1 |
Hipot Return Lead ,1.8m(1102) |
×1 |
×1 |
- |
Ground Bond Test Lead 30A ,1.5m(1103) |
- |
- |
×1 |
Ground Bond Return Lead 30A ,1.5m(1104) |
- |
- |
×1 |
*product specifications are subject to change without notice. |